Specifications and Layout
■
Component dimensions and weights ................................... 109
■
QuantStudio
®
3D Digital PCR Instrument layout and connections ......... 110
■
Instrument clearances ................................................ 110
■
Electrical specifications ............................................... 111
■
Environmental requirements .......................................... 112
■
Network requirements ............................................... 113
Component dimensions and weights
Ensure that the installation site bench space can accommodate the system component
dimensions and weight.
Table 6 Uncrated dimensions and weights
Component Height
Length
(depth)
Width Weight
QuantStudio
®
3D Instrument
21 cm
(8.3 in)
23.3 cm
(9.2 in)
13.5 cm
(5.3 in)
2.4 kg
(5.3 lbs)
QuantStudio
®
3D Digital PCR Chip
Loader
11 cm
(4.3 in)
27 cm
(10.6 in)
[1]
14 cm
(5.5 in)
3.0 kg
(6.1 lbs)
ProFlex
™
2x Flat PCR System
(assembled)
27.2 cm
(10.7 in)
56.5 cm
(22 in)
33.0 cm
(13 in)
19.6 kg
(43.2 lbs)
Dual Flat Block GeneAmp
®
PCR
System 9700 (assembled)
26.0 cm
(10.0 in)
41.0 cm
(16.0 in)
28.0 cm
(11.0 in)
11.8 kg
(26.0 lbs)
[1]
Length increases to 38 cm (14.9 in) when open.